Manual Water Meter Test Bench Double Series

Manual Water Meter Test Bench Double Series

Size:DN15~50mm
Accuracy:≤0.2%
Warranty Period:1 year
Send Inquiry

Hebei Shanghong Meters Technology Co., Ltd. is one of the leading manufacturers and suppliers of manual water meter test bench double series in China. If you're going to wholesale high quality manual water meter test bench double series, welcome to get pricelist and quotation from our factory.

 

 

Manual Water Meter Test Bench

 

Manual-Water-Meter-Test-Bench1

 

Manual Water Meter Test Bench Description

 

Product Name

Double Series Manual Water Meter Test Bench

DN

DN15~25mm

DN15~50mm
Test Quantity
DN15*10 DN20*8 DN25*7 per row (Detection quantity can be customized)
DN15*20 DN20*18 DN25*16  DN32*1 DN40*1 DN50*1 (Detection quantity can be customized)
Water Pump

380V 3KW

380V 7.5KW+3KW
Size
4000mm*2000mm*2440mm
4500mm*1950mm*2800mm
Verification method Volumetric method

Working Temperature

0~50℃

Air Pressure 0.35~0.6Mpa

Motor

380V, three-phase, 50Hz

Accuracy

≤0.2%

Warranty Period

1 year

 

Manual water meter Test Bench – Working Principle

 

Working Principle: The test medium is drawn by a water pump and then stabilized by a pressure tank to reduce fluctuations and pulsations in the flow. It is delivered into the test pipeline, passes through the meter under test and the flow indication device, and then enters the corresponding measuring instrument. The readings of the meter under test are compared with those of the standard device to determine the measurement error of the water meter.

 

Manual water meter Test Bench – Composition

 

The manual water meter test bench of S.H. Meters mainly consists of a workbench, pressure tank, water pump, graduated cylinder, water tank, filter, glass rotor flow meter, hydraulic system, pressure gauge, temperature gauge, and various valves

 

Manual Water Meter Test Bench – Operating Instructions

 

1. Place the water meter to be tested on the test bench and secure it. Then, open the inlet valve to pressurize the pipeline and the water meter, ensuring that the meter is properly clamped and there are no leaks.

 

2. Open all flow control valves in sequence to purge the air from the pipeline and the water meter under test. Once the air is completely removed, close all flow control valves.

 

3. Record the initial reading of the water meter, open the flow control valve corresponding to the desired flow, and adjust it to the target flow rate. During the process, carefully observe the level in the measuring device. When the level reaches the specified position, quickly close the flow control valve

 

4 . Record the final reading of the water meter and the actual reading of the measuring device. Drain the water from the measuring device and proceed to test the next flow point. After completing all tests, calculate the error of the water meter based on the collected data

 